A vibrotactile wristband to guide suturing practice

This project built a silicone wristband with twelve vibration motors that signals a direction (up, down, left or right) to someone practicing a suture. It was tested with 18 medical students. After a brief calibration, lateral directions were recognized in 83.3% of attempts, and the best set of patterns reached 80% overall accuracy. No participant reported skin irritation. The wristband has not yet been used during a real suturing task, and the artificial-intelligence system that should decide when to vibrate is proposed but not implemented.
Context
Suturing requires coordinating both hands, following a trajectory with the needle and dosing the force. Bench models give no immediate feedback, and the simulators that do are expensive and scarce. The proposal sits in between: a low-cost device on the wrist that does not get in the way of the instruments and communicates corrections through touch.
The wristband
The structure is a sheet of HC-25 silicone (Shore A) with the motors embedded and a PLA housing that holds the ESP32. The microcontroller drives twelve independent channels through PWM and hosts its own web server, from which the evaluator triggers the patterns during tests.
The silicone was vacuum degassed (−26 inHg for 20 min) to remove bubbles, which would make its response nonlinear. Tensile testing according to ASTM D412 gave a Young's modulus of 0.491 ± 0.012 MPa and an elongation at break of 177 ± 34%, enough to adapt to wrists of different sizes without permanent deformation.
The motors operate between 140 and 220 Hz, the range to which the Pacinian corpuscles of the skin are most sensitive. A direction is encoded by activating a row or a column of motors in sequence, so that the vibration seems to travel across the wrist.
The thickness of the silicone
The motor, silicone and skin assembly was modeled as a mass-spring-damper system, in which thickness sets the stiffness of the layer. A thick layer damps the stimulus; a thin one transmits well but is fragile. According to the paper, the 1.5 to 2.0 mm range gave the highest transmission within the operating range.
Transmissibility as a function of frequency for thicknesses from 1.0 to 3.0 mm, taken from the project presentation. The band marks the operating range of the actuators
Results with users
The test had two phases. First, participants identified the direction without knowing what the patterns were like. They were then shown each pattern and repeated the test. Calibration raised lateral accuracy from 72.7% to 83.3%. In the figure, down is the hardest direction: it stays below 60% even after calibration.
Accuracy per direction before (dashed line) and after (solid line) learning the patterns
A nine-item Likert questionnaire (Cronbach's α > 0.80) assessed perception, comfort and weight. Fifteen of the 18 participants found the weight appropriate and felt no discomfort, and all 18 denied irritation. Clarity of the vibration was the weakest point: ten agreed that they perceived it clearly and five disagreed.
Responses of the 18 participants to questions Q1 to Q9
What is missing
The test measured whether the vibration can be distinguished, not whether it helps to suture better. Trials during real suturing, a comparison with training without the wristband and a measurement of retention are missing.
The artificial-intelligence component is a design: a camera with markers on the pad would estimate the needle trajectory and choose the pattern. In the tests, the patterns were triggered by a person.
Proposed closed loop: computer vision detects the trajectory error and the wristband communicates it through vibration
The paper does not report the data from the thickness characterization: how many samples were tested, with which sensor, or the measured values. The curves in the figure above have the ideal shape of a single-degree-of-freedom model. In addition, an earlier version of the text refers to linear resonant actuators (LRA) rather than eccentric rotating masses (ERM). The difference matters, because in an ERM the frequency rises with voltage together with the amplitude and cannot be set separately. It would be worth clarifying which actuator was used and measuring transmission with an accelerometer on the skin.
How it fits in Robiolab
This project takes haptics, one of the lab's central interests, to medical education. It complements the ferrofluid haptic actuator, which seeks a softer and more localized stimulus, and the characterization of the impedance between a soft actuator and a finger, which provides the kind of measurement missing here. The idea of informing the user through vibration already appeared in the myoelectric prosthesis. It leaves a working device, a user-evaluation protocol and a first set of tested patterns. The work was presented at SAMYRAD 2026 in Seville.